Colour medical imaging

The same colours on every screen that matters

Dental, pathology, endoscopy, ophthalmology and dermatology teams make decisions from colour photographs and images. ViziCAL calibrates each display to sRGB and measures the result, so the colour you see is the colour in the image.

Colour is part of the image

A shade, a stain, the colour of tissue or skin: in these images colour carries clinical information, and it is judged on a screen.

No two displays show colour alike

Models differ from one another, and a display changes as it ages. The same photograph can look different on the next screen along, or on the same screen next year.

One measured target for every screen

ViziCAL calibrates each display to sRGB, the colour space most cameras and images use, and measures the result, so every calibrated workstation shows the image the same way.

CIE 1931 chromaticity diagram. The coloured area is every colour people can see. The triangle is sRGB (IEC 61966-2-1), the target ViziCAL calibrates colour displays to: red primary at x 0.64, y 0.33; green at x 0.30, y 0.60; blue at x 0.15, y 0.06; and the D65 white point at x 0.3127, y 0.3290.0.00.10.20.30.40.50.60.70.80.00.10.20.30.40.50.60.70.80.9xy460480520540560580600620700RedGreenBlueD65
sRGB (IEC 61966-2-1), ViziCAL’s colour targetD65 white pointCIE 1931 chromaticity, wavelengths in nm. Colours outside sRGB are shown as the nearest a screen can display.

What is measured

Calibrated to sRGB, and proved

After calibrating, ViziCAL measures the display again against sRGB (IEC 61966-2-1) and its D65 white point, and passes or fails it on these results.

Average ΔE2000
How far the measured colours are from their sRGB targets on average. Passes at 2.0 or less.
Maximum ΔE2000
The largest difference of any one colour. Passes at 3.0 or less.
Gamut coverage
The share of sRGB the calibrated display covers. Passes at 99% or more, or when each primary is within tolerance.
Native gamut coverage
The display’s own range before correction, for choosing and comparing displays.

A real calibration

A real display, calibrated to sRGB

A DELL U3223QE calibrated in ViziCAL 2.4.0 with the ViziCAL X-Rite Photometer. The chart shows where each check colour was measured against its sRGB target, and the results passed every limit.

Average ΔE2000
0.5
limit 2.0
Maximum ΔE2000
1.9
limit 3.0
sRGB coverage
99.5%
limit 99%
The ViziCAL app Colour Gamut chart after a colour calibration: the CIE 1931 chromaticity diagram with the sRGB and DCI-P3 gamuts, the D65 white point and the red, green and blue primaries, and each check colour’s measured point beside its sRGB target and the small tolerance outline around it.
A DELL U3223QE calibrated to sRGB in the ViziCAL app (version 2.4.0) with the ViziCAL X-Rite Photometer: average ΔE2000 0.5, maximum 1.9, sRGB gamut coverage 99.5%.

Who it’s for

For every team that reads colour from a screen

Dental practices and groups

Intraoral and extraoral photographs and shade records, reviewed in the surgery, in the lab and between clinicians. Dental X-ray displays can be calibrated to the DIN 6868-157 dental presets on the same software.

Pathology and histology

Stained sections, digital slides and specimen photographs, where the colour of a stain is part of what is being read.

Endoscopy

Still images and recordings reviewed on Windows workstations after the procedure, where the colour of tissue matters.

Ophthalmology and optometry

Fundus photographs and other retinal images, compared over time and shared with colleagues.

Dermatology

Clinical and dermoscopic photographs of the skin, compared between visits and between clinicians.

Every group that reviews colour images

Wound care, plastic and reconstructive surgery, medical photography departments and telehealth teams: any professional medical or dental group that makes decisions from colour images on screen.

How it works

Calibrating a colour display

  1. Choose Colour Medical Photography for the computer in ViziCAL’s settings.
  2. Place the ViziCAL Photometer on the screen. ViziCAL measures a series of colour patches and the grey scale of each colour channel.
  3. ViziCAL builds a colour profile for the display and makes it the display’s Windows colour profile, so programs with colour management use it straight away.
  4. It checks the result twice: as colour-managed programs show colours, and as every other program shows them.
  5. You get the results, a PDF calibration report and the display’s colour history in the app.

Good to know

  • Every program gets the corrected greys and white point; programs with colour management, such as most photo viewers and web browsers, also get the full colour correction. The report shows both.
  • Colour is measured with the ViziCAL Photometer, the same photometer used for radiology displays.
  • A display is calibrated for colour or for radiology (DICOM), not both. Radiology reading displays in the same practice use ViziCAL’s DICOM calibration.
  • Colour calibration is part of ViziCAL: the same workstation licence and photometer, with no separate colour product to buy.
  • Dental X-ray displays can use the DIN 6868-157 RC5 (dental diagnosis) and RC6 (dental treatment) presets. See display standards.

See colour calibration on your own displays

Tell us which images you review and on how many workstations, and we’ll show you ViziCAL in a demo.