Standards

Display standards and acceptance criteria by region

Every display is calibrated to the DICOM Grayscale Standard Display Function and checked against the preset you choose for it. Here is each preset ViziCAL ships, the guidance it is based on and the limits it checks.

ACR–AAPM

Mainly used in
United States
Based on
ACR–AAPM–SIIM Technical Standard for Electronic Practice of Medical Imaging (revision 2017)
Published by
American College of Radiology (ACR) and American Association of Physicists in Medicine (AAPM)
ACR–AAPM presets: the limit each value must meet to pass. A dash means the preset sets no limit.
PresetGSDF error% at mostL′maxcd/m² at leastL′mincd/m² at leastLuminance ratioat leastWhite point (Δu′v′ from D65)at mostAmbient (Lamb ÷ Lmin)at mostRoom illuminancelux at most
ACR AAPM General Reporting1035012500.010.2575
ACR AAPM Mammography104201.22500.010.2575
ACR AAPM Secondary Monitors202500.82500.020.2575

RANZCR

Mainly used in
Australia and New Zealand
Based on
RANZCR Standards of Practice for Diagnostic and Interventional Radiology (version 11.0, 2019)
Published by
RANZCR
RANZCR presets: the limit each value must meet to pass. A dash means the preset sets no limit.
PresetGSDF error% at mostL′maxcd/m² at leastL′mincd/m² at leastLuminance ratioat leastUniformity% at mostWhite point (Δu′v′ from D65)at mostAmbient (Lamb ÷ Lmin)at mostRoom illuminancelux at most
RANZCR Reporting (V11)103501250300.020.2575
RANZCR Mammography Reporting (V11)104501.2250300.020.2575
RANZCR Secondary Displays (V11)202500.8250300.020.2575

RCR

Mainly used in
United Kingdom
Based on
RCR Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices, third edition (February 2019)
Published by
Royal College of Radiologists (RCR)
RCR presets: the limit each value must meet to pass. A dash means the preset sets no limit.
PresetGSDF error% at mostL′maxcd/m² at leastL′mincd/m² at leastLuminance ratioat leastWhite point (Δu′v′ from D65)at mostRoom illuminancelux at most
UK RCR Reporting (2019)1035012500.0275
UK RCR Secondary Review (2019)202500.82500.0275
UK RCR Reporting Mammography (2019)104001.22500.0275

DIN 6868-157

Mainly used in
Germany
Based on
DIN 6868-157:2022-01, Image quality assurance in diagnostic X-ray departments, Part 157
Published by
DIN, Deutsches Institut für Normung

Each preset is one of the standard’s room classes (RC1 to RC6).

DIN 6868-157 presets: the limit each value must meet to pass. A dash means the preset sets no limit.
PresetGSDF error% at mostL′maxcd/m² at leastL′mincd/m² at leastLuminance ratioat leastAmbient (Lamb ÷ Lmin)at mostRoom illuminancelux at most
DIN RC1 Reading Room1025012500.90950
DIN RC2 Exam Room1015011000.909100
DIN RC3 Procedure Room1515011000.909500
DIN RC4 Treatment Room1515011000.9091,000
DIN RC5 Dental Diagnosis1020011000.909100
DIN RC6 Dental Treatment1030011000.9091,000

JESRA

Mainly used in
Japan
Based on
JESRA X-0093-2005, Quality Assurance (QA) Guideline for Medical Imaging Display Systems (August 2005)
Published by
Japan Medical Imaging and Radiological Systems Industries Association (JIRA)
JESRA presets: the limit each value must meet to pass. A dash means the preset sets no limit.
PresetGSDF error% at mostL′maxcd/m² at leastL′mincd/m² at leastLuminance ratioat leastWhite point (Δu′v′ from D65)at mostRoom illuminancelux at most
JESRA Grade 11517012500.011,000
JESRA Grade 23010011000.011,000

Universal

Mainly used in
Several regions: Australia, Germany, Japan, New Zealand, the United Kingdom and the United States
Based on
ViziCAL’s own combination of the ACR–AAPM, DIN, JESRA, RANZCR and RCR presets above, taking the highest performance criteria of each
Universal presets: the limit each value must meet to pass. A dash means the preset sets no limit.
PresetGSDF error% at mostL′maxcd/m² at leastL′mincd/m² at leastLuminance ratioat leastUniformity% at mostWhite point (Δu′v′ from D65)at mostAmbient (Lamb ÷ Lmin)at mostRoom illuminancelux at most
Universal General Reporting103501250no limit0.010.2575
Universal Mammography Reporting104501.2250300.010.2575
Universal Clinical Specialist Displays202500.8250300.020.2575

Colour displays and your own limits

  • Displays used for colour medical photography are checked against sRGB (IEC 61966-2-1:1999). See colour calibration for medical imaging.
  • You can also set your own limits as user-defined criteria. Reports then show your own title, and the display can’t be described as conforming to a published standard.

What ViziCAL measures and checks

The app measures each display with the ViziCAL Photometer, or its built-in photometer, and passes or fails it against the limits of its preset. The same figures reach the web portal with each calibration.

GSDF error
How far the display’s contrast response is from the DICOM Grayscale Standard Display Function.
L′max and L′min
The brightest white and darkest black, including the room light reflected off the screen.
Luminance ratio
L′max divided by L′min: how much brighter the display’s white is than its black.
Uniformity
How evenly bright the display is across the screen, as a percentage difference.
White point
How close the display’s white is to D65. It counts towards the result when D65 colour calibration is switched on; otherwise a miss is shown as an advisory.
Ambient light
Room light reflected off the screen (Lamb, compared with the display’s black level) and room light falling on the screen (illuminance, in lux).

Your standard isn’t listed? Tell us which one you work to.

DICOM, ACR, AAPM, SIIM, RANZCR, RCR, DIN, JIRA (JESRA) and IEC are referenced for their published standards and guidance. These references don’t imply certification or endorsement of ViziCAL by those organisations.